It may seem silly buy I have had offices think that they were trying different combinations when in reality they were using the same interface/sensor combo for testing.
#Installing dexis platinum hardware serial
This can be done by tracking serial numbers.

If you still get a white image in the different room/tube-head than we know we have factored out the tube-head being an issue.

There a a few different factors that can result in white images or no images. When you are testing the sensors are you testing by putting the sensor in someone’s mouth or are you simply laying something metal on the sensor and putting the tube head right over the sensor?.Do you have more than one x-ray tube-head?.Is this happening on multiple computers?.Have you tried a sensor that you know for a fact is working correctly, such as a sensor from another Dr’s office?.Did the white image issue start happening all at the same time?.When you say you are using a box do you mean the DexUSB box or the Dexbox? (The DexUSB box is small, all white, and says DexUSB on the back of the box.).We have re-installed the program, tried a different box and use 2 diff sensors with the same result. Our system just started picking up white images. "We are using a dexis dental sensor with a box. If you are currently having issues and are not using powered USB hubs I suggest making this your first move not only in troubleshooting a faulty sensor but also prolonging the life of your current and future digital x-ray sensors. *Note: We highly recommend using powered USB hubs with not just dexis dental sensors but ALL digital x-ray sensors. The goal of this troubleshooting method is to determine if the issue with a Dexis dental sensor is being caused by: